Compact 2 V Cryocooled Programmable Josephson Voltage Standard for Direct On-Site Quantum Voltage Comparisons
Resources
2 resources available
-
Fig 2 Comparison results at 2V
CSV -
4086_README.txt
TXT
Find Related Datasets
Search by Tags
Click any tag below to search for similar datasets
Complete Metadata
| @type | dcat:Dataset |
|---|---|
| accessLevel | public |
| bureauCode |
[ "006:55" ] |
| contactPoint |
{ "fn": "Alain Rufenacht", "hasEmail": "mailto:alain.rufenacht@nist.gov" } |
| description | This paper evaluates the performance of a new type of 2 V traveling cryocooled Programmable Josephson Voltage Standard for performing on-site verification of quantum voltage standards. Direct comparison with a 10 V system showed agreement of 8 parts in 10^11 at 2 V. This result shows that, despite the lower voltage, the direct Josephson-to-Josephson system comparison method provides a two orders-of-magnitude improvement over traveling Zener standards for verifying Josephson voltage standards. |
| distribution |
[ { "title": "Fig 2 Comparison results at 2V", "mediaType": "text/csv", "description": "Figure 2 presents the voltage measured as a function of duration including three different dither current values and two bias configurations", "downloadURL": "https://data.nist.gov/od/ds/mds2-4086/Fig%202%20Comparison%20results%20at%202V.csv" }, { "title": "4086_README.txt", "mediaType": "text/plain", "description": "Read me file, with description of the dataset.", "downloadURL": "https://data.nist.gov/od/ds/mds2-4086/4086_README.txt" } ] |
| identifier | ark:/88434/mds2-4086 |
| issued | 2026-03-17 |
| keyword |
[ "Josephson junctions", "Measurement uncertainty", "Standards", "Superconducting integrated circuits", "Voltage measurement." ] |
| language |
[ "en" ] |
| license | https://www.nist.gov/open/license |
| modified | 2026-02-02 00:00:00 |
| programCode |
[ "006:045" ] |
| publisher |
{ "name": "National Institute of Standards and Technology", "@type": "org:Organization" } |
| theme |
[ "Metrology:Electrical/electromagnetic metrology" ] |
| title | Compact 2 V Cryocooled Programmable Josephson Voltage Standard for Direct On-Site Quantum Voltage Comparisons |