Reinstatement of the NIST Field Strength Probe Calibration Service

Published by National Institute of Standards and Technology | National Institute of Standards and Technology | Catalog Last Checked: August 02, 2025 at 02:54 PM | Dataset Last Updated: April 05, 2022
After recently completing a five year renovation, the NIST anechoic chamber for radio-frequency field strength measurements was returned to service. Tests of field probe transfer functions can now be performed from 10 kHz to 40 GHz. To validate that the chamber and measurement process maintained continuity, measurements of NIST field probes in the new chamber were compared with measurements in the chamber before renovation and in another test facility at NIST. An uncertainty analysis was carried out on the new measurement process.